BioMajesty® JCA-BM6010/C
The JCA-BM6010/C is designed to increase the performance of medium-sized laboratories. It is also the ideal system for speciality laboratories. The throughput of up to 1.200 tests/hour, 43 reagent and 84 sample positions are guarantee for the flexibility in everyday use. It handles a full menue of photometric and immunoturbidimetric assays as well as Na, K, Cl determination by indirect ISE methods.
Optimized Workflows
Special emphasis has been spent on the software to combine highest user friendliness with optimally secured results. A special feature is the integrated on-board hemolysis function for optimized HbA1c determination. Dynamic range extension makes additional dilutions superfluous and a dedicated STAT port allows for immediate STAT processing. Clot detection and liquid level sensor technology ensures confidence in results.
Patient Friendly
The analyzer works with very small sample volumes, making it the perfect instrument in pediatric and geriatric settings.
Economic
The JCA-BM6010/C is the system with the smallest footprint in its class hence saving precious laboratory space. DiaSys supplies ready-to-use reagents well known for their excellent on-board stability with the instrument. The result in conjunction with extremely low reagent consumption is cost efficiency on the highest order.
Technical Specifications
| System type | Floor stand random access clinical chemistry analyzer |
| Total throughput | 1200 tests/hour |
| Sample tray STT | 84 positions for routine and priority samples with built-in barcode reader |
| Refrigerated sample tray CTT | 61 positions for controls and calibrators |
| Sample types | Serum, plasma, CSF, urine, whole blood |
| Reagent trays | Two refrigerated trays with 50 positions each for reagent containers with; Built-in barcode reader |
| Sample dispensing volume | 1-25 µL (0.1 µL step) |
| Reagent pipetting | 5 to 300 µL |
| STAT-analytics | STAT port enables easy STAT sample determination |
| Photometry throughput | 800 tests/hour |
| ISE throughput (Na, K, CI) | 600 tests/hour |
| Sample throughput rate | Maximum 800 samples/hour |
| Measuring principle | Colorimetry (Rate/End Point); Turbidimetric Immunoassay; ISE: simultaneous 3 item measurement of Na, K, CI |
| Barcode identification | Automatic barcode scan for reagents and samples |
| Calibration | Linear, non-linear, multi point |
| Sample tubes/cups | Most commonly used blood collection tubes and sample cups |
| Sample dilution | Dilution ratio: 2 to 300 times |
| Sensors | Liquid-level sensor, clot sensor and crash sensor |
| Reagent dispensing method | 2 probe system capable to dispense 3 reagent components |
| Reagent tray standard setting | Total 50 containers per tray with 40 mL |
| Reagent container capacity | 20 and 40 mL with special adapters for 20 mL containers |
| Reaction disk | Rotating reaction carousel with turntable and 231 reusable plastic cuvettes |
| Reaction times and temperature | 3,4, 5,10,15 minutes at 37 ± 0.1 °C; Extended reaction time: 21 or 31 minutes |
| Reflex testing capability | Yes |
| Photometry | 14 wavelengths: 340, 410, 451, 478, 505,545,571, 596, 658, 694,751, 805, 845, 884 nm |
| Automatic maintenance | Automatic startup and shutdown by weekly timer |
| Required pure water | 20 Liter/hour; water purity: ≤ 1 ps/cm |
| Ambient temperature/humidity | Temperature:18 °C to 30 °C/humidity: 40 % to 70 % |
| Power supply | AC 220 V +10 %, 50/60 Hz; 2.6 kVA |
| Dimensions | 122 cm (W) x 85 cm (D) x 110.8 cm (H) |
| Weight | Approximately 450 kg |