Trusted Performance
The respons® 920 achieves a throughput of 200 tests/hour (360 tests/hour with ISE) regardless of whether mono- or 2-component tests are performed.
Customer-oriented
High on-board capacity of 30 different methods in mono or twin containers and 30 sample positions offer the necessary flexibility for everyday use. Several sample positions are available for STAT analytics. A comprehensive range of high-quality clinical chemistry and immunoturbidimetric DiaSys tests rounds off our parameter portfolio.
Ready to use reagents, adapter-free one-grip loading of reagent containers, automated bar code scan of sample and reagent as well as the software with intuitive user interface facilitates routine work significantly.
Economic
Not only is the reagent consumption low, but DiaSys reagents are also well-known for their long-term stability. Broad measuring ranges minimize the need for retesting.
Technical Specifications
| System type | Bench top random access clinical chemistry analyzer |
| Throughput | 200 tests/hour with a cycle time of 18 seconds for mono- and 2 component tests 360 tests/hour with ISE |
| Sample tray | 30 positions for bar coded patient samples including STAT positions, 9 positions for blanks, calibrators, controls, samples without barcode and ISE solutions |
| Sample types | Serum, plasma, urine, CSF, whole blood |
| Reagent tray | 30 different methods in bar coded mono - or twin - containers for adapter-free one-grip loading, refrigerated |
| Sample volume | 2 – 70 μL |
| Reagent pipetting volume | Reagent 1: 50 – 300 μL Reagent 2: 10 – 200 μL |
| STAT-analytics | Several sample positions available |
| Bar code identification | Automatic bar code scan for reagents and samples |
| Measuring principle | Colorimetry (Rate/End Point); Turbidimetric Immunoassay; Direct Potentiometry (ISE: K, Na, Li, Cl) |
| Calibration | Linear, non-linear, multi point |
| Sample tubes/cups | Most commonly used blood collection tubes and sample cups |
| Sample dilution | Dilution ratio: 2 to 150 times |
| Reaction temperature | 37 ± 0.2 °C |
| Reaction unit | Temperature controlled rotating tray with 45 reusable quartz glass cuvettes (37 ± 0.2 °C) |
| Photometry | 8 wavelengths: 340, 405, 450, 505, 546, 578, 660 and 700 nm (mono- and bichromatic) |
| Photometric linearity and resolution | Linearity: 0 – 2.5 OD; resolution: 0.0001 OD |
| Water consumption | Up to 7.5 liters per hour |
| System interface | Analyzer - PC: USB connectivity bi-directional; CPU: Pentium IV or higher |
| Power source | AC 220 V ± 10 %, 50 ± 1 Hz or AC 110 V ± 10 %, 60 ± 1 Hz ; 600 VA (excluding PC/printer/monitor) |
| Dimensions | 81 cm (W) x 70 cm (D) x 60 cm (H) |
| Weight | Approximately 110 kg |