Trusted Performance

The respons® 920 achieves a throughput of 200 tests/hour (360 tests/hour with ISE) regardless of whether mono- or 2-component tests are performed.

Customer-oriented

High on-board capacity of 30 different methods in mono or twin containers and 30 sample positions offer the necessary flexibility for everyday use. Several sample positions are available for STAT analytics. A comprehensive range of high-quality clinical chemistry and immunoturbidimetric DiaSys tests rounds off our parameter portfolio.

Ready to use reagents, adapter-free one-grip loading of reagent containers, automated bar code scan of sample and reagent as well as the software with intuitive user interface facilitates routine work significantly.

Economic

Not only is the reagent consumption low, but DiaSys reagents are also well-known for their long-term stability. Broad measuring ranges minimize the need for retesting.

Technical Specifications

System type Bench top random access clinical chemistry analyzer
Throughput 200 tests/hour with a cycle time of 18 seconds for mono- and 2 component tests
360 tests/hour with ISE
Sample tray 30 positions for bar coded patient samples including STAT positions, 9 positions for blanks, calibrators, controls, samples without barcode and ISE solutions
Sample types Serum, plasma, urine, CSF, whole blood
Reagent tray 30 different methods in bar coded mono - or twin - containers for adapter-free one-grip loading, refrigerated
Sample volume 2 – 70 μL
Reagent pipetting volume Reagent 1: 50 – 300 μL
Reagent 2: 10 – 200 μL
STAT-analytics Several sample positions available
Bar code identification Automatic bar code scan for reagents and samples
Measuring principle Colorimetry (Rate/End Point); Turbidimetric Immunoassay; Direct Potentiometry (ISE: K, Na, Li, Cl)
Calibration Linear, non-linear, multi point
Sample tubes/cups Most commonly used blood collection tubes and sample cups
Sample dilution Dilution ratio: 2 to 150 times
Reaction temperature 37 ± 0.2 °C
Reaction unit Temperature controlled rotating tray with 45 reusable quartz glass cuvettes (37 ± 0.2 °C)
Photometry 8 wavelengths: 340, 405, 450, 505, 546, 578, 660 and 700 nm (mono- and bichromatic)
Photometric linearity and resolution Linearity: 0 – 2.5 OD; resolution: 0.0001 OD
Water consumption Up to 7.5 liters per hour
System interface Analyzer - PC: USB connectivity bi-directional; CPU: Pentium IV or higher
Power source AC 220 V ± 10 %, 50 ± 1 Hz or AC 110 V ± 10 %, 60 ± 1 Hz ; 600 VA (excluding PC/printer/monitor)
Dimensions 81 cm (W) x 70 cm (D) x 60 cm (H)
Weight Approximately 110 kg

Reagent Kits for respons® Analyzers