General

  • Precise photometer for measurement of enzymes, substrates and immunoturbidimetric tests
  • End point, fixed time and kinetic absorbance with immunoturbidimetric measurments
  • High quality interference filters for minimum band widthand accurate results7 filters (340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 578 nm and 630 nm) + 1 free position
  • Dynamic temperature control by Peltier elements
  • Up to 112 programmable tests and memory for 3200 patient results
  • Touch screen, large LCD display
Light source Quartz-halogen lamp 12V/20W
Wavelength range Automatic by 8 positions filter-wheel; 7 standard filters: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 578 nm, 630 nm 1 position free for optional filter; Half bandwidth: < 8 nm at 340 nm
Photometric range 0.0000 to 3.0000 ABS, resolution: 0.0001 ABS, stability: 0.005 ABS / hour
Flowcell Stainless steel with quartz window, measuring volume: 32 μL, optical path: 10 mm, aspiration volume programmable: 200 – 2000 μL, carryover: < 1.0%, temperature control: 25, 30 and 37°C
Analysis method End point, with or without reagent blank, end point, with sample blank, with or without reagent blank, kinetic with or without linearity check, two-point kinetic, fixed time, with or without reagent blank, absorbance, turbidimetry, linear and non-linear calibration
Parameter settings Method, wavelength, temperature, reagent blank y/n, sample blank y/n, lag time, measurement time, reaction time, absorbance limit, aspiration volume, standards, linearity check, unit for results
Memory 112 test profiles, 3200 sample results
Temperature control By means of Peltier elements 25°C, 30°C, 37°C (± 0.2°C)
Input Touch screen
Display Large LCD display
Print Built-in thermal printer
Interface RS-232 serial port
Operation environment Temperature: 15°C – 30°C, Humidity: ≤ 20 – 80%
Power requirement AC 220 V, 110 V (± 10%), 50 Hz (± 2%); AC 40 V ± 10%, 50/60 Hz
Dimension 470 mm (H) x 440 mm (W) x 230 mm (D)
Weight 8.5 kg