A step ahead

respons®910 is a compact and fully-automated bench top analyzer for maximum efficiency. During its development, special attention was paid to easy handling and optimized work flow thus leading to the ideal clinical chemistry system for many laboratories.

To use this service, your consent is required for data protection reasons. Click "Accept and Play" to consent to the processing of your personal data by YouTube LLC. For more information, please see our „Privacy Policy“.

To use this service, your consent is required for data protection reasons. Click "Accept and Play" to consent to the processing of your personal data by YouTube LLC. For more information, please see our „Privacy Policy“.

Versatile and flexible

High on-board capacity of 30 different methods in mono- or twin-containers and 30 sample positions offers the necessary flexibility for everyday use. The integrated STAT port allows easy loading of emergency samples at any time. The wide range of high-quality clinical chemistry and immunoturbidimetric tests manufactured by DiaSys rounds off the respons®910 system.

Easy to use

Ready-to-use reagents, adapter free one-grip loading of reagent containers, integrated bar code reader for sample and reagent as well as the intuitive software alleviates daily workload significantly.

Competitive and unique

Minimized service and maintenance costs, low reagent consumption and high on board-stability of DiaSys system reagents in combination with highest result safety add up for a highly competitive system. Further features like LIS-connectivity, clot detection and crash sensor make respons®910 unique in its class.

The newest software version 4.8.9.6 is available for respons®910 analyzers. DiaSys highly recommends to install this newest version on all respons®910 systems in the market to maintain the analyzer up to date.

Download Software (please login)

System type Bench top clinical chemistry analyzer
Throughput 150 tests/hour for mono-reagents; 100 tests/hour for 2 component reagents
Combined reagent/sample tray 30 reagent positions plus 30 patient sample positions, easy removable tray for storage in refrigerator
Sample types Serum, plasma, whole blood, urine
Sample volume 2 – 30 μL
Reagent pipetting volume Reagent 1: 120 – 250 μL Reagent 2: 10 – 50 μL
STAT-analytics 2 sample positions for loading of emergency samples at any time
Ion measurement Enzymatic tests for Cl, K, Na, Li
Bar code identification Automatic bar code scan for reagents and samples
Measuring principle Colorimetry (Rate/End Point); Turbidimetric Immunoassay
Calibration Linear, non-linear, multi-point
Sample tubes/cups Primary tubes of 5, 7 and 10 mL and sample cups; special mini cup adapter for application areas where sample material is limited (pediatrics, veterinary medicine)
Reagent on-board capacity 30 different methods in bar coded mono- or twin-containers for adapter-free one-grip loading
Reaction temperature 37 ± 0.2 °C
Reaction unit Temperature controlled heated rotor with 105 disposable plastic cuvettes (37 ± 0.2 °C)
Photometry 12 wavelengths: 340, 380, 405, 450, 480, 508, 546, 570, 600, 660, 700 and 800 nm (mono- and bichromatic)
Photometric linearity and resolution 0 – 3.0 OD; resolution: 0.0001 OD
Water consumption < 1 liter per hour
LIS connectivity Given
System interface Analyzer - PC: USB connectivity bi-directional; CPU: Pentium IV or higher
Power source/power consumption AC 110/220 V, 60/50 Hz; 300 VA excluding PC/printer/monitor
Dimensions 60 cm (W) x 67 cm (D) x 60 cm (H)
Weight Approximately 60 kg