A step ahead
respons®910 is a compact and fully-automated bench top analyzer for maximum efficiency. During its development, special attention was paid to easy handling and optimized work flow thus leading to the ideal clinical chemistry system for many laboratories.
General
Versatile and flexible
High on-board capacity of 30 different methods in mono- or twin-containers and 30 sample positions offers the necessary flexibility for everyday use. The integrated STAT port allows easy loading of emergency samples at any time. The wide range of high-quality clinical chemistry and immunoturbidimetric tests manufactured by DiaSys rounds off the respons®910 system.
Easy to use
Ready-to-use reagents, adapter free one-grip loading of reagent containers, integrated bar code reader for sample and reagent as well as the intuitive software alleviates daily workload significantly.
Competitive and unique
Minimized service and maintenance costs, low reagent consumption and high on board-stability of DiaSys system reagents in combination with highest result safety add up for a highly competitive system. Further features like LIS-connectivity, clot detection and crash sensor make respons®910 unique in its class.
The newest software version 4.8.9.6 is available for respons®910 analyzers. DiaSys highly recommends to install this newest version on all respons®910 systems in the market to maintain the analyzer up to date.
Download Software (please login)
System type | Bench top clinical chemistry analyzer |
Throughput | 150 tests/hour for mono-reagents; 100 tests/hour for 2 component reagents |
Combined reagent/sample tray | 30 reagent positions plus 30 patient sample positions, easy removable tray for storage in refrigerator |
Sample types | Serum, plasma, whole blood, urine |
Sample volume | 2 – 30 μL |
Reagent pipetting volume | Reagent 1: 120 – 250 μL Reagent 2: 10 – 50 μL |
STAT-analytics | 2 sample positions for loading of emergency samples at any time |
Ion measurement | Enzymatic tests for Cl, K, Na, Li |
Bar code identification | Automatic bar code scan for reagents and samples |
Measuring principle | Colorimetry (Rate/End Point); Turbidimetric Immunoassay |
Calibration | Linear, non-linear, multi-point |
Sample tubes/cups | Primary tubes of 5, 7 and 10 mL and sample cups; special mini cup adapter for application areas where sample material is limited (pediatrics, veterinary medicine) |
Reagent on-board capacity | 30 different methods in bar coded mono- or twin-containers for adapter-free one-grip loading |
Reaction temperature | 37 ± 0.2 °C |
Reaction unit | Temperature controlled heated rotor with 105 disposable plastic cuvettes (37 ± 0.2 °C) |
Photometry | 12 wavelengths: 340, 380, 405, 450, 480, 508, 546, 570, 600, 660, 700 and 800 nm (mono- and bichromatic) |
Photometric linearity and resolution | 0 – 3.0 OD; resolution: 0.0001 OD |
Water consumption | < 1 liter per hour |
LIS connectivity | Given |
System interface | Analyzer - PC: USB connectivity bi-directional; CPU: Pentium IV or higher |
Power source/power consumption | AC 110/220 V, 60/50 Hz; 300 VA excluding PC/printer/monitor |
Dimensions | 60 cm (W) x 67 cm (D) x 60 cm (H) |
Weight | Approximately 60 kg |